Catalog Number 199721001S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a revision surgery occurred on (b)(6) 2024 due to screw displacement and migration.In the surgery, it was confirmed that sfx broke off as well as the reported screws.The deviated screw was replaced.The fenestrated screws were added for the extension on upper vertebrae.All of the set screw were replaced.Sfx was performed.The surgery was completed successfully.Patient outcome was stable the original surgery occurred on (b)(6) 2024 and was for a spinal fusion (l5/s) (l4-sai fusion using verse), for pseudarthrosis on (b)(6) 2024.Patient was stable.Original procedure was completed successfully with no surgical delay.It was noted post-operatively on an unknown date that the screws had migrated.This report involves one (1) 5.5 exp verse unitized set scr this is report 5 of 6 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b3: unknown event date d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E1: initial reporter is j&j company representative h3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d9: device returned.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device and pictures provided.Visual analysis of the returned sample revealed that 5.5 exp verse unitized set scr was chipped at the hexlobular internal driving feature.Fragments were not received and the overall surface of the device was noted with signs of usage which could have been result of implantation and explantation.A dimensional inspection for the sfx,5.5,ti, med, size a5 was not performed due to post manufacturing damage.The chipped condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the sfx,5.5,ti, med, size a5 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): a manufacturing record evaluation was performed for the finished device product code:199721001s.Lot no: l20340.It was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on:29-nov-2023.Manufacturing site: jabil le locle.Expiry date: 31-oct-2028.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE UNITIZED SET SCR
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key19519144
Report Number1526439-2024-01658
Device Sequence Number6171563
Product Code NKB
UDI-Device Identifier10705034466149
UDI-Public(01)10705034466149
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K142185
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number199721001S
Device Lot NumberL20340
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/16/2024
Supplement Date Received by Manufacturer06/19/2024
Initial Report FDA Received Date06/12/2024
Supplement Report FDA Received Date07/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.5 EXP VERSE UNITIZED SET SCR; UNK - LOCKING/SET SCREWS: EXPEDIUM SFX; UNKNOWN SCREWS
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-