| Catalog Number |
383312 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd saf-t-intima w/prn yel 24ga x.75in had foreign matter.On (b)(6) 2024, bed 1 of the obstetrics and gynecology inpatient department was prepared for use by patients.When the needle cap was opened, it was found that there was a foreign body on the tip of the needle, which was like lint.Replace the indwelling needle with a new one immediately.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record review was completed by our quality engineer team for provided material number 383312 and lot number 2146189.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|
Search Alerts/Recalls
|