| Model Number |
130° RADIOLUCENT TARGETING ARM |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 05/16/2024, it was reported by a sales representative via (b)(4) that an 0826-000 impactor pad dethreaded the metal coupling where it attaches to the 125° jig.As a result, the impactor pad could not be threaded into the 130° jig.While using the 130° jig, the sleeves' retaining button was sticking, preventing them from holding in place.This was discovered during a case with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user-applied excessive force during the tightening process.
|
| |
|
Search Alerts/Recalls
|