| Model Number |
28164LBA |
| Medical Device Problem Codes |
Solder Joint Fracture (2324); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device will be forwarded to the manufacturing site for investigation.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted unsolicited.This event is filed under internal complaint id (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a right frontal neuro endoscopic arachnoid cyst fenestration on (b)(6) 2024, the scissors broke.They were able to complete the procedure by using a biopsy forcep since they used the scissor for majority of the procedure until it stopped working properly.There was no patient injury reported.Patient had ct scan to ensure scissors did not break inside of the patient, and nothing was found per the ct result.
|
| |
|
Additional Manufacturer Narrative
|
|
Changes made from the initial report: updated the udi number in section d4.Updated section d9 to yes, and entered the device returned date to the manufacturer.Per investigation by the manufacturing site: during the investigation, it was found that the drawbar was probably torn due to overloading, and the cutting edge also showed clear damage caused by improper use of the scissors.We therefore assume that this was clearly an user error.This event is filed under internal complaint id (b)(4).
|
| |
|