• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 125° RADIOLUCENT TARGETING ARM; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 125° RADIOLUCENT TARGETING ARM; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number 125° RADIOLUCENT TARGETING ARM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
On 05/16/2024, it was reported by a sales representative via (b)(4) that an 0826-000 impactor pad de-threaded the metal coupling where it attaches to the 125° jig.As a result, the impactor pad could not be threaded into the 130° jig.While using the 130° jig, the sleeves' retaining button was sticking, preventing them from holding in place.This was discovered during a case with no patient harm.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user-applied excessive force during the tightening process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
125° RADIOLUCENT TARGETING ARM
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19527611
Report Number1220246-2024-05668
Device Sequence Number20720287
Product Code HSB
UDI-Device Identifier00848665028652
UDI-Public00848665028652
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K021008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number125° RADIOLUCENT TARGETING ARM
Device Catalogue Number1267-100
Device Lot Number172549
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2024
Supplement Date Received by Manufacturer06/06/2025
Initial Report FDA Received Date06/13/2024
Supplement Report FDA Received Date06/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-