W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER Back to Search Results
Catalog Number DSF1833
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 05/14/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The following was reported to gore: on (b)(6) 2024 a gore® dryseal flex introducer sheath (dsf) was being used an accessory with non-gore devices to clear a blood clot.At some point in the procedure, significant bleeding was noticed through the one-way valve.The decision was made to replace the device with a different dsf.However, the new dsf also had significant bleeding through the one-way valve.The patient lost a significant amount of blood (approximately 4-5 units).The patient underwent a transfusion and is stable.Because the primary procedure was not conducted with gore device, the fsa was not present at this event.The fsa reported that it was indicated to him that in both cases, the valve leakage occurred after a device had been advanced through valve.However, the hospital risk management indicated that this information was unknown, and would not be provided.It is unknown how the procedure was completed.
 
Additional Manufacturer Narrative
A.1.Patient identifier was requested from hospital but was not provided.H.6.Type of investigation code c21: investigation pending follow-up conversations with hospital risk management, and product history review, and engineering evaluation.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
H.6.Investigation conclusions code d15: there is insufficient information available for gore to reasonably draw conclusions related to aspects of the event, therefore conclusion code ¿d15: cause not established¿ is being used.Insufficient information may include limited or missing relevant medical records, involvement of multiple implanted devices (including non-gore devices) in the field of treatment, patient non-compliance, and/or a general lack of available detail or specificity related to an adverse event and/or device.H.6.Type of investigation code b17: the hospital would not confirm whether they had retain or discarded the device.H.6.Type of investigation code b22: a review of the manufacturing records for the device could not be conducted because the serial/lot number remains unknown.H.6.Investigation conclusions code d16 updated to d15.H.6.Investigation findings code c21 update to c20 and c0703.H.6.Type of investigation code b21 updated to b13, b17, and b22.
 
Additional Manufacturer Narrative
The event is this case is the same event documented in user facility medwatch #mw5155105.
 
Additional Manufacturer Narrative
H.6.Investigation findings code c19: the device history record was reviewed to ensure that each manufacturing and inspection operation was performed and indicated the product met creganna medical specifications and procedures.H.6.Type of investigation code b22 corrected to b14.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® DRYSEAL FLEX INTRODUCER SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
bryce wargin
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key19527804
Report Number3007284313-2024-03308
Device Sequence Number18036043
Product Code DYB
UDI-Device Identifier00733132630042
UDI-Public00733132630042
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K160254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDSF1833
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2024
Supplement Date Received by ManufacturerNot provided
Not provided
Not provided
Initial Report FDA Received Date06/13/2024
Supplement Report FDA Received Date08/05/2024
09/03/2024
09/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age34 YR
Patient SexFemale
-
-