|
The sample was evaluated under ambient light conditions.The sample arrived with a circled area on the blue layer where there is an incomplete daisy pattern.The area measures approximately 1.2cm.The area was viewed under magnification.The area can be described as a flatter bond point area where the daisy pattern is not defined.The white layer is not affected.The complaint is undergoing further investigation by the manufacturing site, a follow up report will be submitted upon conclusion of investigation.This product incident is documented in the o&m halyard, inc.Complaint database and identified as complaint comp (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an o&m halyard, inc.Product is defective or has caused serious injury.
|
|
A complaint sample was evaluated under ambient light conditions.The sample arrived with a circled area on the blue layer where there is an incomplete daisy pattern.The area measures approximately 1.2cm.The area was viewed under magnification.The area can be described as a flatter bond point area where the daisy pattern is not defined.The white layer is not affected.The device history record for product was reviewed.Quality records showed the attributes and visual checks on the rolls were documented as pass.There were no documented issues.According to quality records, this material met specification guidelines.There can be times where polymer or particles can stick to the rolls.These small particles or debris of foreign material can transfer to the sheet, which can cause a hole, make an indention, or cause flattened areas on the sheet.Product testing is conducted utilizing a lot acceptance system and aql limits.Some of the process tools utilized to minimize issues include online web inspection detection devices, process settings with established run settings, center lining control plans, an automatic process monitoring system, and statistical process controls.Results of the process and product testing are closely monitored.Customer feedback is reviewed which utilizes metrics such as cpm (complaints per million) and trend analysis to monitor the effectiveness of process and quality control.No actions are taken at this time, the observation is considered a visual defect that does not affect the strength and durability of the fabric.This product incident is documented in the o&m halyard, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an o&m halyard, inc.Product is defective or has caused serious injury.
|