It was reported that prior to patient contact for an unspecified procedure, a hair like fiber found in the packaging of a safe-t-j fixed core wire guide.The procedure was completed as normal with a new wire.As reported, a patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
|
Summary of event: it was reported that prior to patient contact for an unspecified procedure, a hair like fiber found in the packaging of a safe-t-j fixed core wire guide.The procedure was completed as normal with a new wire.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.A hair-like fiber was found inside the already opened returned pouch.A document-based investigation evaluation was performed.A review of the device history record found a relevant non-conformance on one device; however, all affected product was scrapped prior to release.A review of complaint history found no additional complaints for this lot number.The device instructions for use (ifu) states, ¿supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened and undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured out of specification.However, there is no evidence of additional non-conforming devices in-house or in the field.Although one relevant non-conformance was noted on the complaint lot, all non-conforming product was scrapped, there are 100% inspections in place to capture this non-conformance, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot-related complaints have been received from the field.Therefore, it was concluded that there is no evidence that additional non-conforming product exists in house or in the field.Based on the information provided and the results of the investigation, cook has concluded that the cause of this failure was a manufacturing and quality control deficiency.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|