• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW, T10 FULL THREAD, 2.7MM/L22MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW, T10 FULL THREAD, 2.7MM/L22MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 657022S
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during orif (left) performed on (b)(6)2024, the locking failure occurred during use of the locking screw with the variax clavicle plate.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidence were provided.More detailed information about the complaint event as well as the affected device must be available to determine the root cause of the complaint event.Related the reported failure mode regarding the locking of the variax2 screw it can be mentioned that as part of our post market surveillance activities a potential non-conformity report (nc) was initiated to address similar events related to the variaax2 locking feature.The deep investigation of the nc revealed that the application of over torque during insertion was the root cause of the event.This leads to the damage of the smart lock technology below the head.As a reminder, the operative technique clearly states that the proper use of the screw should prevent this deformation from happening: caution: with the use of variable speed power systems, the surgeon should initially reduce the power to the lowest setting.Final tightening of the screw should be performed by hand to avoid damaging the screw-plate interface¿.However, although no product related issue could be detected a preventive action (capa) was nonetheless opened to make the design more robust against a potential over tightening by the user.A review of the device history for the reported lot did not indicate any abnormalities.No deviation from the specifications was noted.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.If any further substantial information or the material is provided, the investigation report will be updated.
 
Event or Problem Description
It was reported that during orif (left) performed on (b)(6) 2024, the locking failure occurred during use of the locking screw with the variax clavicle plate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, T10 FULL THREAD, 2.7MM/L22MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19533389
Report Number0009610622-2024-00285
Device Sequence Number7483619
Product Code HWC
UDI-Device Identifier07613327071306
UDI-Public07613327071306
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K180500
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/10/2024
Device Catalogue Number657022S
Device Lot Number1000363998
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/16/2024
Supplement Date Received by Manufacturer07/16/2024
Initial Report FDA Received Date06/13/2024
Supplement Report FDA Received Date08/06/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/24/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
-
-