| Catalog Number |
0600560 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 11/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that three weeks and one day post a chronic catheter placement, the catheter was allegedly ruptured.It was further reported that the connection of the catheter came loose and allegedly broke off.Reportedly, the catheter was repaired.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.Two photos were provided for review.The photo shows a completely fractured luer adapter from the crimp collar and a crack was noted to the crimp collar.Therefore, the investigation is confirmed for the reported broke off issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2026), g3, h6 (method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that three weeks and one day post a chronic catheter placement, the catheter was allegedly ruptured.It was further reported that the connection of the catheter came loose and allegedly broke off.Reportedly, the catheter was repaired.There was no reported patient injury.
|
| |
|