Model Number STRIPPER SEMITENDINOSUS, 7MM
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Event description: it was reported by an arthrex employee that there is rust on the laser markings.The reported event was confirmed as the visual evaluation revealed that the device shows signs of metal surface oxidation.Further the laser marks are slightly faded.Based off the information provided, the most likely cause of the reported failure is wear and tear.
 
Event or Problem Description
It was reported by an arthrex employee that there is rust on the laser markings.There was no harm or adverse event for patient, operator or third party reported.No further information received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRIPPER SEMITENDINOSUS, 7MM
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19535525
Report Number1220246-2024-05760
Device Sequence Number6171285
Product Code LXH
UDI-Device Identifier00888867006904
UDI-Public00888867006904
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSTRIPPER SEMITENDINOSUS, 7MM
Device Catalogue NumberAR-1278L
Device Lot Number011712
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2021
Initial Report FDA Received Date06/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/27/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-