|
ARTHREX, INC. INSTRUMENT CASE, UNIVERS VAULTLOCK GLEN; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
|
Back to Search Results |
|
| Model Number |
INSTRUMENT CASE, UNIVERS VAULTLOCK GLEN |
| Medical Device Problem Code |
Pitted (1460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.Three ar-9217c sets (no batch/sn info present) were received for investigation.Visual inspection identified scuff markings across the outer coating of the sets that were consistent with wear and tear damage, as well as peeling of the coating within the trays.This is most likely the result of repeated cleaning cycles, and it is unknown how many cleaning cycles the trays have been subjected to.
|
| |
|
Event or Problem Description
|
|
It was reported that the following sets are chipped and need to be replaced; ar-9217c, ar-9400c4, ar-9225c, ar-9226abc, ar-9226ubc, ar-9226cc, ar-9501-hc2, ar-9501-hc1, ar-9615c-1, ar-9615c-2.
|
| |
|
Search Alerts/Recalls
|
|
|