• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE Back to Search Results
Catalog Number EN2914
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
 
Event or Problem Description
According to the available information, the device was explanted and replaced due to a tubing leakage.
 
Additional Manufacturer Narrative
Titan touch pump and cylinders 1 and 2 were received for evaluation.No functional abnormalities were noted with the pump.No functional abnormalities were noted with cylinder 1 or cylinder 2.The information received indicated the device had a tubing leakage, but because no functional abnormalities were noted with the returned components, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
 
Event or Problem Description
According to the available information, the device was explanted and replaced due to a tubing leakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key19536693
Report Number2125050-2024-00892
Device Sequence Number20376340
Product Code FHW
UDI-Device Identifier05708932539098
UDI-Public05708932539098
Combination Product (Y/N)N
Initial Reporter CountryKR
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberEN2914
Device Lot Number8458977_EN29141022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/14/2024
Supplement Report FDA Received Date08/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
-
-