OLYMPUS WINTER & IBE GMBH OBTURATOR, 24 FR.; HF-RESECTION ELECTRODE Back to Search Results
Model Number A22081A
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code Internal Organ Perforation (1987)
Date of Event 05/18/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.This report is related to (b)(6).
 
Event or Problem Description
It was reported that during an unknown procedure, there was a bladder perforation.The physician felt as if it was due to an electrical problem with the resectoscope.The patient had to undergo a laparotomy due to the perforation.Additional information was requested multiple times, but not received.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was evaluated by olympus, and no abnormalities were observed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than 1 year since the subject device was manufactured.Based on the results of the investigation, no device abnormalities were noted.Therefore, the root cause of the event could not be determined.This supplemental report includes additional information received from the customer.B5 updated accordingly.A correction has been made to d1, d2, d4, g2, and g4 from the initial medwatch.An update has been made to d9 and h3.Also, information has been added to b2, h4, and h6.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported that the incident occurred during a bladder resection using bipolar energy.The bladder perforation occurred when the surgeon approached the patient¿s obturator nerve.The patient was not injured by electric current and/or shock.The generator did not display an error message.The generator was programmed to a ¿classic¿ bipolar setting recommended by olympus.Due to the bladder perforation, an emergency laparotomy was performed.This resulted in a 1-hour procedural delay.The surgeon reported the incident to the hospital¿s biomedical team.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OBTURATOR, 24 FR.
Common Device Name
HF-RESECTION ELECTRODE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19538767
Report Number9610773-2024-01475
Device Sequence Number20416403
Product Code GCP
UDI-Device Identifier04042761021159
UDI-Public04042761021159
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K931994
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA22081A
Device Lot Number21502
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2024
Supplement Date Received by Manufacturer07/24/2024
Initial Report FDA Received Date06/14/2024
Supplement Report FDA Received Date07/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HF CABLE.; OUTER SHEATH.; WORKING ELEMENT.
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-