• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CASHEL SABERX RADIANZ; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL SABERX RADIANZ; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number R51007004L
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/21/2024
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the balloon of a 7mm x 4cm saberx radianz percutaneous transluminal angioplasty (pta) balloon catheter ruptured at nominal pressure.A non-cordis balloon catheter was used to complete the procedure.There were no reported injuries to the patient.This was during a procedure to treat a 90% stenosed superficial femoral artery (sfa) lesion which had mild calcification and mild tortuosity.The device was stored and prepped per the instructions for use (ifu) and there were no difficulties removing any of the sterile packaging components.The device was able to maintain negative pressure during preparation.The device was able to be removed easily from the patient and remained in one piece during its removal.The device will not be returned for evaluation.
 
Additional Manufacturer Narrative
A review of the manufacturing documentation associated with lot 82279078 presented no issues during the manufacturing process that can be related to the reported event.Additional information is pending and will be submitted within 30 days upon receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections have been updated accordingly: b4, g3, g6, h1, h2, h3, h6, and h10 complaint conclusion: as reported, the balloon of a 7mm x 4cm saberx radianz percutaneous transluminal angioplasty (pta) balloon catheter ruptured at nominal pressure.A non-cordis balloon catheter was used to complete the procedure.There were no reported injuries to the patient.This was during a procedure to treat a 90% stenosed superficial femoral artery (sfa) lesion which had mild calcification and mild tortuosity.The device was stored and prepped per the instructions for use (ifu) and there were no difficulties removing any of the sterile packaging components.The device was able to maintain negative pressure during preparation.The device was able to be removed easily from the patient and remained in one piece during its removal.The device will not be returned for evaluation.The product was not returned for analysis.A product history record (phr) review of lot 82279078 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst-at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics of mild calcification with stenosis likely contributed to the rupture reported.Calcification is known to damage balloon material; it is likely damage occurred during crossing or in the attempt to inflate the balloon at the lesion.However, without the return of the device for analysis it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use, ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.Neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABERX RADIANZ
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, co. tipperary
E25 R C92
EI  E25 RC92
Manufacturer (Section G)
CORDIS CASHEL
cahir road
cashel, co. tipperary
Manufacturer Contact
karla castro
coyol free zone, building b25, ,  , costa rica
el coyol alajuela bld 25 
7863138372
MDR Report Key19539522
Report Number3007635982-2024-00156
Device Sequence Number20341165
Product Code LIT
UDI-Device Identifier20705032081808
UDI-Public20705032081808
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date06/30/2026
Device Catalogue NumberR51007004L
Device Lot Number82279078
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2024
Supplement Date Received by Manufacturer06/18/2024
Initial Report FDA Received Date06/14/2024
Supplement Report FDA Received Date06/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NEW BALLOON CATHETER (NON-CORDIS).
Patient SexUnknown
-
-