MAUDE Adverse Event Report: EMBECTA MEDICAL I LLC - HOLDREGE, NE SYRINGE 1.0ML 31GA 8MM UFII 10BAG 500CS; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EMBECTA MEDICAL I LLC - HOLDREGE, NE SYRINGE 1.0ML 31GA 8MM UFII 10BAG 500CS; SYRINGE, PISTON Back to Search Results
Model Number 328418
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Consumer reported the needle bends during injection.Then it clogs and can't get full dose.Denied reuse of needles.Does not prefill syringes.Dc.Consumer reported when removed shield from the needle on 1 syringe from this box the needle was bent.Informed caller to never straighten needles.She does this.Dc.Lot # 5306833.Catalog# 328418.Date of event unknown.Sample status awaiting samples the 2 from above and will send back an unopened bag.Dc.
 
Additional Manufacturer Narrative
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st.Complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 1.0ML 31GA 8MM UFII 10BAG 500CS
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
EMBECTA MEDICAL I LLC - HOLDREGE, NE
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key19547248
Report Number3024508819-2024-00121
Device Sequence Number4733971
Product Code FMF
UDI-Device Identifier00382908418034
UDI-Public00382908418034
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K190054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number328418
Device Lot Number5306833
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/17/2024
Supplement Report FDA Received Date09/03/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/06/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-