| Model Number |
N/A |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: cat# 30123203 lot# 66484387 32mm i.D.Size c high wall liner cat# 30123203 lot# 65889222 32mm i.D.Size c high wall liner cat# 30123204 lot# 65588016 32mm i.D.Size d high wall liner cat# 30123204 lot# 65781505 32mm i.D.Size d high wall liner g2: foreign ¿ japan.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial hip procedure, the shell was inserted and the liner could not be inserted.The attempt was made to insert a second liner, which could also not be inserted.A second shell was inserted.The attempt was made to insert two new liners and both could not be inserted.A third liner was able to inserted into the shell to complete the surgery.The surgeon reported that the shell was not stable during liner insertion due to fragile bone quality.There was a 10-15 minute delay to surgery.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: g3; h2; h3; h6 the second shell was not returned as it was implanted.A review of the device history records identified no related deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.Event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial hip procedure, the shell was inserted and the liner could not be inserted.The attempt was made to insert a second liner, which could also not be inserted.A second shell was inserted.The attempt was made to insert two new liners and both could not be inserted.A third liner was able to inserted into the shell to complete the surgery.The surgeon reported that the shell was not stable during liner insertion due to fragile bone quality.The surgeon used screws to fix the unstable cup.There was a 10-15 minute delay to surgery.There was no known impact or consequences to the patient.Attempts have been made and no further information has been provided.
|
| |
|