• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. PEGGED GLENOID TRIL & DRILL GUIDE CADDY; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. PEGGED GLENOID TRIL & DRILL GUIDE CADDY; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number PEGGED GLENOID TRIL & DRILL GUIDE CADDY
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.Visual inspection identified corrosion spots inside the caddy.Based off the information provided, the most likely cause for the reported failure can be attributed to user error of the device due to improper detergent/cleaning process used.
 
Event or Problem Description
It was reported by the agency that the following devices are corroded: ar-9231-01pp qty 1; ar-9231-02pp qty 1; ar-9231-03pp qty 1; ar-9231-04pp qty 1; ar-9217cpp qty 2; ar-9225ck qty 1; ar-9225cp qty 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEGGED GLENOID TRIL & DRILL GUIDE CADDY
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19548313
Report Number1220246-2024-05881
Device Sequence Number6716227
Product Code MDM
UDI-Device Identifier00888867137769
UDI-Public00888867137769
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPEGGED GLENOID TRIL & DRILL GUIDE CADDY
Device Catalogue NumberAR-9225CP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2021
Initial Report FDA Received Date06/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-