• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOT DR UNIT HNDCONT POWERMAX ELITE; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOT DR UNIT HNDCONT POWERMAX ELITE; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200872
Medical Device Problem Codes Mechanical Problem (1384); Overheating of Device (1437); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation revealed a blade stall error and overheating of the housing.Further evaluation revealed the drive fork was stiff when rotated.The root cause was associated with a mechanical component failure.Factors that could have contributed to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This could be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Several corrective actions were taken to reduce the occurrence of this failure.No further actions are required at this time.
 
Event or Problem Description
It was reported that the mdu was broken.No case reported; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
H10 h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation revealed a blade stall error and overheating of the housing.Further evaluation revealed the drive fork was stiff when rotated.The root cause was associated with a mechanical component failure.Factors that could have contributed to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This could be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Several corrective actions were taken to reduce the occurrence of this failure.No further actions are required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOT DR UNIT HNDCONT POWERMAX ELITE
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19551569
Report Number1643264-2024-00420
Device Sequence Number20515716
Product Code HAB
UDI-Device Identifier03596010618894
UDI-Public03596010618894
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72200872
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2024
Supplement Date Received by Manufacturer06/21/2024
Initial Report FDA Received Date06/17/2024
Supplement Report FDA Received Date06/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-