| Catalog Number |
72200872 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Overheating of Device (1437); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation revealed a blade stall error and overheating of the housing.Further evaluation revealed the drive fork was stiff when rotated.The root cause was associated with a mechanical component failure.Factors that could have contributed to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This could be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Several corrective actions were taken to reduce the occurrence of this failure.No further actions are required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that the mdu was broken.No case reported; therefore, there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation revealed a blade stall error and overheating of the housing.Further evaluation revealed the drive fork was stiff when rotated.The root cause was associated with a mechanical component failure.Factors that could have contributed to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This could be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Several corrective actions were taken to reduce the occurrence of this failure.No further actions are required at this time.
|
| |
|
Search Alerts/Recalls
|