This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.New information added to the following fields: h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the customer's reportable malfunction was confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the insulation tip's damage occurred due to improper handling, application of excessive force, mechanical impact like fall, shock or similar stress, thermal mechanical overload.The event can be detected and prevented by handling the device in accordance with the following instructions for use: chapter 4.1 ¿inspection and testing¿ of the instructions for use (ifu) regarding the proper reprocessing of the affected device.Prior to using medical devices, please ensure they are in perfect technical condition.See also the warnings in the instructions for use (ifu).4 before use warning: infection control risk: properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product: visually inspect the product.Make sure that it has: no corrosion no dents no scratches ceramic insulation at distal end: visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning: risk of injury: impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
|