| Catalog Number |
0250070441 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was damaged/compromised insulation.
|
| |
|
Event or Problem Description
|
|
It was reported that there was damaged/compromised insulation.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: power too high or activated too long.Probable root cause: 1.Generator power malfunction.2.Material/design error.3.Conductive irrigant used.4.Manufacturing / assembly error.5.User error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Search Alerts/Recalls
|