|
Updated field: d4 (udi#), d9, g3, g6, h2, h3, h4, h6, h11.The mayo-hegar nh 6 (120135) was returned for evaluation.The device history record (dhr) was reviewed, and no anomalies related to the reported issue were identified.The reported complaint was confirmed: the returned mayo-hegar nhs were received in used condition with staining and wear on the surface.One of the needle holders had misaligned ratchets, allowing the lock to come apart when struck against a table.The misalignment may be due to wear or environmental damage.
|