MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OR3O DM XLPE INSERT 28/44; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OR3O DM XLPE INSERT 28/44; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 71358219
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 09/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a right hip revision surgery on (b)(6) 2022 in which smith & nephew were implanted; patient sustained a dislocation of the internal right hip prosthesis that required re-revision surgery on (b)(6) 2022.Current patient's health status is unknown.No other complications were reported.
 
Additional Manufacturer Narrative
Internal reference number: (b)(4).H10: this complaint was opened by smith+nephew to document a patient complication identified through the review of clinical evidence from post market clinical data collection activities that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
H3, h6: the devices were not returned for evaluation; therefore, a device analysis could not be performed.The devices' batch numbers were not provided, thus, an evaluation of the manufacturing records could not be performed.For the insert and the liner, a review of complaint history of the previous 12 months revealed a similar event for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.For the femoral head, a review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed that dislocation has been identified in warnings and precautions, that may result from improper selection of the neck length and cup, stem positioning, muscle looseness and/or malpositioning of components.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.At this time, we have no evidence to conclude that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, patient condition, postoperative care, size selected or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H6 (medical device problem code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OR3O DM XLPE INSERT 28/44
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19553696
Report Number1020279-2024-01174
Device Sequence Number11945488
Product Code LPH
UDI-Device Identifier00885556712382
UDI-Public00885556712382
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K191002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71358219
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2024
Supplement Date Received by Manufacturer08/01/2024
Initial Report FDA Received Date06/17/2024
Supplement Report FDA Received Date08/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71342812/OXNM FEM HD 28MM +12; 71358205/OR3O DUAL MOBILITY LINER 44/56
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-