Model Number CA190RC1
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/21/2024
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, during percutaneous ablation procedure, the procedure was ceased.The radiologist was concerned that the therapeutic range was not being met with antenna 1.He water tested the first antenna and according to the radiologist, he was not convinced that the antenna worked, and staff had to open another antenna to ensure delivery in predictable margins.The water test was done after removing the second antenna from the patient and the radiologist was not convinced that the equipment was working.The generator was working fine with different cable and a new antenna.The representative was contacted and was very helpful with troubleshooting but the anesthetist wanted the procedure to be ceased and rebooked.The patient will be rebooked for the procedure.
 
Additional Manufacturer Narrative
D10 concomitant products: ca15l2, ca15l2 15cm rein percutaneous antenna x1, (lot #s24ag006) cagen1, cagen1 ablation generator x1, (sn: (b)(6) ca15l2, ca15l2 15cm rein percutaneous antenna x1, (lot #s24ag006).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPRINT
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer (Section G)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19554569
Report Number3004962788-2024-00075
Device Sequence Number12464054
Product Code GEI
UDI-Device Identifier10884521190764
UDI-Public10884521190764
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K133821
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCA190RC1
Device Catalogue NumberCA190RC1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2024
Initial Report FDA Received Date06/17/2024
Date Device Manufactured09/06/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight75 KG
-
-