|
It was reported that during an arthroscopy, the bioraptor knotless anchor got broken while tapping into the bone, the broken pieces were removed from the meniscus with the help of grasper, it was necessary to drill an additional hole.The procedure was completed using a back-up device.There was a delay less than 30 minutes and no further complications were reported.
|
|
Internal complaint reference (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material specifications found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A clinical review states that based on the limited information provided, definitive contributing clinical factors cannot be concluded; however, ifu provides warnings, precautions and instructions which address breakage; therefore, a user technique and/or procedural variance cannot be completely ruled out as a possible contributing factor.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|