• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MINERVA SURGICAL INC MINERVA ENDOMETRIAL ABLATION SYSTEM, DISPOSABLE HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MINERVA SURGICAL INC MINERVA ENDOMETRIAL ABLATION SYSTEM, DISPOSABLE HANDPIECE Back to Search Results
Catalog Number MIN9770
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Bowel Burn (1756); Uterine Perforation (2121)
Date of Event 05/29/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The disposable handpiece used in this case was not returned and therefore a failure analysis of the complaint device could not be performed.The lot number was not provided by customer so device history review was not performed, the handpieces met all qa specifications prior to being released, including adequate sterilization.Uterine/bowel perforations are known complications of an endometrial ablation procedure.Complications associated with perforations are addressed in the warning and caution sections of the operator's manual and instructions for use, including the statements: use caution not to perforate the uterine wall when sounding, dilating, or inserting the disposable handpiece.Activation of the minerva disposable handpiece in the setting of a uterine perforation is likely to result in serious patient injury.Excessive force applied during placement of the disposable handpiece may result in tissue injury, including perforation.Although designed to detect a perforation of the uterine wall, this test is an indicator only, and it might not detect all perforations.Clinical judgment must always be used.Post-treatment, any patient-reporting signs/symptoms that could indicate a serious complication, e.G., bowel injury, should be thoroughly evaluated without delay.As with all endometrial ablation procedures, serious injury or death can occur, including but not limited to, infection and injury to adjacent tissue, such as bowel, bladder, vagina, vulva, and/or perineum.
 
Event or Problem Description
It was reported, a diagnostic hysteroscopy, post ablation, was performed showing evidence of a fundal uterine perforation, the patient was discharged shortly after.Later that same day the patient began to experience lower abdominal pain and on 5/29 the patient went to the emergency department.Laparoscopy was performed and a uterine perforation was confirmed with an area of unintended thermal effect on the bowel.A small bowel resection and end to end anastomosis was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINERVA ENDOMETRIAL ABLATION SYSTEM, DISPOSABLE HANDPIECE
Common Device Name
MINERVA ENDOMETRIAL ABLATION SYSTEM
Manufacturer (Section D)
MINERVA SURGICAL INC
4255 burton drive
santa clara CA 95054
Manufacturer (Section G)
MINERVA SURGICAL INC
4255 burton drive
santa clara CA 95054
Manufacturer Contact
michelle becker
4255 burton drive
santa clara, CA 95054
9787601704
MDR Report Key19554878
Report Number3011011193-2024-00012
Device Sequence Number20515749
Product Code MNB
UDI-Device Identifier00862393000162
UDI-Public00862393000162
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P140013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberMIN9770
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2024
Initial Report FDA Received Date06/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient SexFemale
-
-