• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER CREEVY MODEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER CREEVY MODEL Back to Search Results
Catalog Number 01267520
Medical Device Problem Codes Inflation Problem (1310); Failure to Infuse (2340)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the foley catheter balloon could not be inflated or discharged during a pretest.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was available for evaluation and further investigation did not result in any additional findings.Root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be "low latex modulus".However, there was insufficient information to confirm this potential root cause.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: "caution: this product contains natural rubber latex which may cause allergic reactions.Sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities: 3cc balloon: use 5ml sterile water; 5cc balloon: use 10ml sterile water; 15cc balloon: use 20ml sterile water; 20cc balloon: use 25ml sterile water; 30cc balloon: use 35ml sterile water; 40cc balloon: use 45ml sterile water; 75cc balloon: use 80ml sterile water; do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." correction: d, f, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the foley catheter balloon could not be inflated or discharged during a pretest.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER CREEVY MODEL
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19557144
Report Number1018233-2024-03525
Device Sequence Number4812081
Product Code EZC
UDI-Device Identifier00801741018800
UDI-Public(01)00801741018800(11)231007(17)260831(10)NGHV4455
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K910195
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Catalogue Number01267520
Device Lot NumberNGHV4455
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2024
Supplement Date Received by Manufacturer10/01/2024
Initial Report FDA Received Date06/18/2024
Supplement Report FDA Received Date10/04/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-