| Model Number |
BB811 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection showed no outward signs of physical damage or abnormalities.Pressure integrity testing was performed at 3 l/pm with 23 psi (1189 mmhg) of back pressure for 10 minutes.During the pressure integrity testing there were no leaks observed from the device including the fiber bundle.The reason for return was not confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a fusion oxygenator, it was reported that following priming of the system but prior to it being connected to the patient the oxygenator began leaking.The device was replaced.There was no patient involvement, so no adverse effect occurred.Medtronic received additional information that the device itself leaked.The leak was from the arterial line port.
|
| |
|