| Model Number |
R1054655TP |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A ventilator was returned to the manufacturer for routine preventative maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, an error code related to the charging of the internal battery was observed.The device's power management board was replaced to address the issue.Additionally, not related to the issue the flow sensor assembly and motor blower assembly due to tobacco smell.
|
| |
|
Additional Manufacturer Narrative
|
|
On previously submitted report, a ventilator was returned to the manufacturer for routine preventative maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, an error code related to the charging of the internal battery was observed.The device's power management board was replaced to address the issue.Additionally, not related to the issue the flow sensor assembly and motor blower assembly due to tobacco smell.The power management board was evaluated by the manufacturer's product investigation laboratory (pil) and found the power management board functioned as designed and operated without issue or error codes.
|
| |
|