• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. PROFEMUR® NECK VAR/VAL 8DG SHORT COBALT CHROME; HIP COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. PROFEMUR® NECK VAR/VAL 8DG SHORT COBALT CHROME; HIP COMPONENT Back to Search Results
Model Number PHAC1252
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/25/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Allegedly, patient underwent a left-sided total hip arthroplasty on (b)(6) 2020.On or about (b)(6) 2024, the patient reported to the surgeon pain, debilitating lack of mobility, and he recommend revision surgery because the device failed due to corrosion at the neck-stem junction.The modular neck slot and femoral neck distal end has corrosive changes and metal debris consistent with cocr damage.Non revised products: product id: p2lxsd36 prime a-class® xlpe standard / lot no.: 1835920 / qty: 1.Product id: p3sbqd54 prime biofoam quad shell / lot no.: 1834699 / qty: 1.
 
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® NECK VAR/VAL 8DG SHORT COBALT CHROME
Common Device Name
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
rachael wise
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key19559348
Report Number3010536692-2024-00347
Device Sequence Number11945736
Product Code JDI
UDI-Device IdentifierM684PHAC12521
UDI-PublicM684PHAC12521
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K091423
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberPHAC1252
Device Catalogue NumberPHAC1252
Device Lot Number1838673
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/18/2024
Supplement Report FDA Received Date11/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-