| Model Number |
UNIVERS VAULTLOCK GLENOID REAMER, MEDIUM |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.One unpackaged ar-9229grpp, batch 04511638 reamer was received for investigation.Visual inspection identified that the edges of the reamer head were worn.As the device was manufactured in 2016, the observed condition is most likely the result of wear and tear damage accumulated over repeated usage.
|
| |
|
Event or Problem Description
|
|
It was reported the ar-9229grpp, glenoid reamer is dull.
|
| |
|