• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN BELLATEK FLEX ABUTMENT 3.4MM (D) 12MM (H) HEXED; DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN BELLATEK FLEX ABUTMENT 3.4MM (D) 12MM (H) HEXED; DENTAL ABUTMENT Back to Search Results
Catalog Number IF3121
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A2: age and date of birth unknown / not provided.A3: patient sex unknown / not provided.A4: weight unknown / not provided.B3: date of event unknown / not provided.E1: email address and last/given name unknown / not provided.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the crown/abutment screw loosened and will not tighten.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111, 4114 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H10: narrative/data was updated.Zimvie did not receive the reported abutment for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are incorrect techniques used, incorrect torque applied - customer error.Therefore, based on the available information, device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN BELLATEK FLEX ABUTMENT 3.4MM (D) 12MM (H) HEXED
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key19560657
Report Number0001038806-2024-01260
Device Sequence Number20534661
Product Code NHA
UDI-Device Identifier00844868043690
UDI-Public(01)00844868043690(17)991231(10)2120008433
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K183138
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberIF3121
Device Lot Number2120008433
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/18/2024
Supplement Report FDA Received Date11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-