EXACTECH, INC. LOGIC TIBIA IMPLANT PSC INSERT, SZ 4, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-44-4009
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988); Lack of Effect (4065)
Health Effect - Clinical Codes Failure of Implant (1924); Synovitis (2094); Osteolysis (2377); Joint Laxity (4526); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Joint effusion (4849)
Date of Event 03/13/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately 144 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, osteolysis, effusion, synovitis, delamination of the tibial polyethylene, worn patella, loosening and joint laxity.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
D10: no concomitants available.The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, loss of range of motion, and loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6, h11 the reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A review of manufacturing data was unable to be performed as the serial/lot information of the product involved in the event was not available.It was discovered that a bag used in the packaging of our polyethylene inserts had been produced outside of our quality specifications.The use of these non-conforming bags may enable increased oxygen diffusion to the uhmwpe (ultra-high molecular weight polyethylene) insert, resulting in increased oxidation of the material relative to inserts packaged with bags that conform to specifications.For inserts with greater than five years of shelf life, increased oxidation of the inserts can lead to premature polyethylene wear resulting in revision surgery.The non-conforming bags were also used in the packaging of our patella components.An additional hhe and crc were held to assess the risk of the non-conforming packaging specifically related to patella components.Patella components packaged in non-evoh vacuum bags, which have a maximum 5-year shelf-life prior to expiration, have reduced risk of oxidative degradation.Since manufacturing information was not available for the components, it could not be confirmed whether the devices were packaged in non-conforming vacuum bags.Additionally, these devices appear to have been implanted over ten years prior to the reported event.Based on the age of the patient¿s devices and the available information, there is no evidence or information to suggest that the reported loosening is manufacturing-related.A review of complaint history determined that no further action is required as no trends were identified.A definitive root cause was unable to be determined; however, may have resulted from prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted over ten years prior to the reported event.Potential contributions of users and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA IMPLANT PSC INSERT, SZ 4, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19561930
Report Number1038671-2024-01973
Device Sequence Number15331831
Product Code JWH
UDI-Device Identifier10885862173638
UDI-Public10885862173638
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K110547
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/09/2019
Device Catalogue Number02-012-44-4009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2024
Supplement Date Received by Manufacturer03/14/2025
04/15/2026
Initial Report FDA Received Date06/18/2024
Supplement Report FDA Received Date03/17/2025
04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/10/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age86 YR
Patient SexMale
-
-