Catalog Number 212186
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Noise, Audible (3273)
Health Effect - Clinical Code Muscle/Tendon Damage (4532)
Date of Event 05/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Straight saw attachment makes extremely loud noise while executing the cuts with a narrow blade.It is suspected that the oscillation motion is wider than normal.Half way of executing the tibia cut, it was noticed the patellar tendon was damaged by the blade.It had to be stitched.
 
Additional Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Additional Manufacturer Narrative
An event regarding inaccurate resection involving a mako saw attachment was reported.The event was not confirmed because the product was not available for inspection.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other complaints with similar event(s) for the lot referenced.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Event or Problem Description
Straight saw attachment makes extremely loud noise while executing the cuts with a narrow blade.It is suspected that the oscillation motion is wider than normal.Half way of executing the tibia cut, it was noticed the patellar tendon was damaged by the blade.It had to be stitched.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7 DEGREE STRAIGHT SAGITAL SAW
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer (Section G)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer Contact
sanjana talathi
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19562149
Report Number3005985723-2024-00074
Device Sequence Number12682807
Product Code OLO
UDI-Device Identifier00848486032111
UDI-Public00848486032111
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K172219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number212186
Device Lot NumberSN 3521797 / LOT 35010222
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/27/2024
Supplement Date Received by Manufacturer08/16/2024
Initial Report FDA Received Date06/18/2024
Supplement Report FDA Received Date09/10/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age68 YR
Patient SexFemale
-
-