| Catalog Number |
ER8075 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information the device was explanted and replaced due to tubing leakage.
|
| |
|
Additional Manufacturer Narrative
|
|
Reservoir was received for evaluation.A separation was noted on the inlet tubing of the reservoir at the strain relief junction.This was a site of leakage.The surfaces appear to be rough and irregular, indicating stress was exerted.Based on examination of the returned product, it was concluded that the rough and irregular surfaces associated with the separation indicates that sufficient stress was exerted on the inlet tube near the reservoir to separate the site while in-vivo.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|
Event or Problem Description
|
|
According to the available information the device was explanted and replaced due to tubing leakage.
|
| |
|
Search Alerts/Recalls
|