| Model Number |
15 |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report a non-critical issue with their device.Upon evaluation of the customer¿s device, stryker observed that the device had logged an event code in the memory which is related to a potential issue of the device to deliver energy.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.The event code was cleared from the memory of the device and thereafter did not return.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.The cause of the reported issue could not be conclusively determined.
|
| |
|