• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FUSED LIGHT CABLE, 5.0 MM X 274CM (9.0'); LIGHT, SURGICAL, FIBEROPTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FUSED LIGHT CABLE, 5.0 MM X 274CM (9.0'); LIGHT, SURGICAL, FIBEROPTIC Back to Search Results
Model Number FUSED LIGHT CABLE, 5.0 MM X 274CM (9.0')
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The device was not received for evaluation, and no photos of the failure were provided.According to the lit-021 arthrex 950-0031-00 k ifu arthrex light guides, revision k.4.Warnings and precautions.Warning: risk of burns! the optical fibers emit high-energy light at the distal end of the fiber-optic cable.This can cause the temperature of the body tissue to rise to 106 °f (41°c).The complaint allegation was not confirmed.No problem found.
 
Event or Problem Description
On 10/14/2022, it was reported by a sales representative via sems that a ar-3240-5027 fused light cable is getting really hot during procedures.This occurred during an unknown case, with no patient effect reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSED LIGHT CABLE, 5.0 MM X 274CM (9.0')
Common Device Name
LIGHT, SURGICAL, FIBEROPTIC
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19563392
Report Number1220246-2024-06026
Device Sequence Number7550155
Product Code FST
UDI-Device Identifier00888867029408
UDI-Public00888867029408
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K901035
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFUSED LIGHT CABLE, 5.0 MM X 274CM (9.0')
Device Catalogue NumberAR-3240-5027
Device Lot NumberWO150554
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Initial Report FDA Received Date06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/16/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-