MAUDE Adverse Event Report: HILL-ROM CARY NURSE CALL; SYSTEM, COMMUNICATION, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM CARY NURSE CALL; SYSTEM, COMMUNICATION, POWERED Back to Search Results
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The hillrom technician found the code call was not configured correctly for some rooms.The navicare® nurse call¿ (nnc) system is a comprehensive communication and information management system that helps caregivers provide the best possible care to patients in the most efficient manner.The system facilitates communication within a nursing unit.Caregivers have the ability to quickly locate and communicate with a fellow staff member, and patients can easily communicate with caregivers, the nursing station, or directly to their primary caregiver.The hrc technician arranged to configure a secondary console to monitor code calls for the customer to resolve the issue.Although there was no injury associated with this event, if the reported issue were to recur it could lead to injury or death.Therefore hillrom is reporting this complaint.
 
Event or Problem Description
Hillrom received a report from a hillrom technician stating the code blue did not annunciate in the secondary location.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NURSE CALL
Common Device Name
SYSTEM, COMMUNICATION, POWERED
Manufacturer (Section D)
HILL-ROM CARY
1225 crescent green dr., suite 300
cary NC 27518
Manufacturer Contact
estephania cardenas
1225 crescent green dr., suite 300
cary, NC 27518
8129310130
MDR Report Key19563408
Report Number2027454-2024-00044
Device Sequence Number18028559
Product Code ILQ
UDI-Device Identifier00887761985209
UDI-Public00887761985209
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2024
Initial Report FDA Received Date06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-