Model Number SUTURE PASSING WIRE
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when the surgeon attempted to pass the 1.3mm tape, 3 ar-1255-18, suture loops were torn off in quick succession.The surgeon used a new product to finish the case.No adverse effect on patient.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.Visual evaluation identified that all four devices had the nitinol wire loop had broken off from the distal end.The wire loop was found to be frayed as well.Due to the fractured state of the device, functional testing could not be performed.Based off the information provided, the most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used when tensioning the sutures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE PASSING WIRE
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19563625
Report Number1220246-2024-06025
Device Sequence Number7549668
Product Code LXH
UDI-Device Identifier00888867006539
UDI-Public00888867006539
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model NumberSUTURE PASSING WIRE
Device Catalogue NumberAR-1255-18
Device Lot Number11726142
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2021
Initial Report FDA Received Date06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-