• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOFIRE DIAGNOSTICS, LLC BIOFIRE® FILMARRAY® GASTROINTESTINAL (GI) PANEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOFIRE DIAGNOSTICS, LLC BIOFIRE® FILMARRAY® GASTROINTESTINAL (GI) PANEL Back to Search Results
Model Number RFIT-ASY-0116
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: a potential false positive vibrio result on the biofire gi panel was reported after testing a patient stool sample in cary blair media on may 13, 2024.It is unknown if the patient was impacted due to the biofire gi panel.No serious injury or death occurred.Conclusion: the investigation determined that the most likely cause of the discrepant vibrio result was the presence of non-viable organism/nucleic acid in the cary blair transport media.While cary blair medium vials are autoclaved and routinely quality controlled for sterility, non-viable organisms or nucleic acids can remain in the transport medium after the sterilization process.The presence of non-viable organisms and nucleic acid does not compromise the intended use of transport medium, transportation and preservation of clinical stool and rectal specimens; however, the biofire gi panel does not distinguish between nucleic acid from viable or non-viable organisms.The "laboratory precaution" and "limitation" sections of the filmarray gi reagent instruction booklet (https://www.Online-ifu.Com/iti0030) outlines the potential for false positive detections during molecular testing with sterile cary blair transport medium containing detectable levels of non-viable organisms and/or nucleic acid.Importantly, results from the biofire gi panel are intended to be correlated with the clinical history, epidemiological data, and other data available to the clinician evaluating the patient.Detection of organism targets does not imply that the corresponding organisms are infectious or are the causative agents for clinical symptoms.Clinical performance: according to table 11.Biofire gi clinical performance summary of the filmarray gi reagent instruction booklet, the performance claim for vibrio (v.Parahaemolyticus/v.Vulnificus/v.Cholerae) compared to stool culture with standard manual/automated microbiological/biochemical identification showed a specificity of 99.9% (95% ci 99.5-100%).Vibrio was detected in 2/2 false positive specimens using bi-directional sequence analysis.To supplement the results of the prospective clinical study, an evaluation of preselected archived specimens was performed.According to table 17.Biofrie gi panel archived specimen performance data summary, the performance claim for vibrio showed a positive percent agreement (ppa) of 100% (95% ci n/a) and a negative percent agreement (npa) of 100% (95% ci 97.1-100%).Additionally, in table 18.Biofire gi panel performance using contrived specimens, vibrio showed a ppa of 97.4% (95% ci 92.6-99.5%) and a npa of 100% (95% ci 94.0-100%).Includes 64/65 v.Cholerae (five different strains were used in spiking; one specimen spiked near the assay limit of detection was not detected) and 48/50 non-v.Cholerae (four v.Parahaemolyticus strains and one v.Vulnificus strain were used in spiking; two specimens spiked with v.Parahaemolyticus near the assay limit of detection were not detected).
 
Event or Problem Description
Summary: (b)(6) reported a potential false positive vibrio result on the biofire® filmarray® gastrointestinal (gi) panel after testing a patient stool sample.Due to the biofire gi panel result, it is unknown if the patient was impacted.The investigation determined that the most likely cause of the discrepant vibrio result was the presence of non-viable organism/nucleic acid in the cary blair transport media.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFIRE® FILMARRAY® GASTROINTESTINAL (GI) PANEL
Common Device Name
BIOFIRE® FILMARRAY® GASTROINTESTINAL (GI) PANEL
Manufacturer (Section D)
BIOFIRE DIAGNOSTICS, LLC
515 colorow dr
salt lake city UT 84108
Manufacturer (Section G)
BIOFIRE DIAGNOSTICS, LLC
515 colorow dr
salt lake city UT 84108
Manufacturer Contact
515 colorow dr
salt lake city, UT 84108
8017366354
MDR Report Key19563844
Report Number3002773840-2024-00358
Device Sequence Number15197118
Product Code PCH
UDI-Device Identifier00815381020109
UDI-Public00815381020109
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K230404
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2025
Device Model NumberRFIT-ASY-0116
Device Catalogue NumberRFIT-ASY-0116
Device Lot Number0451124
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2024
Initial Report FDA Received Date06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age9 YR
Patient SexMale
-
-