It was reported that on (b)(6) 2024, the patient underwent an unknown surgery.The nail and the drill bit interfered when drilling a burr hole for the blade.The connector and device were retightened, the buttress nut was checked for excessive crimping, and drilling was done again, but interference still occurred.Since the outer cortex was able to be shaved, drilling with a stepped ream f (03.037.022) allowed reaming without interference.The surgery was completed successfully within 30 minutes delay.The surgeon commented as follows: the nail was forcibly inserted without reaming the medullary cavity although the patient's femoral medullary cavity was narrow and bowing was strong.Thus, the nail became flexed and interference between the nail and the drill bit occurred.No further information is available.This report is for one (1)10mm cannulated tapered drill bit.This is report 2 of 2 for complaint (b)(4).
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|