| Model Number |
20001100100011360 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device would not power up.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch reports the evaluation of the device.This supplemental medwatch report also corrects information submitted in the initial medwatch report.Please reference sections d4 model number updated and serial #, d4 catalog number removed, d4 primary udi number updated.Evaluation results: the device was evaluated by zoll medical corporation.The customer's report was not replicated or confirmed.The device was put through extensive testing including self-testing using known good pads and batteries, shock testing, and stress testing without duplicating the customer's report.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|