| Model Number |
SYNERGY RESECTION SHAVER CONSOLE |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023. arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to a failing capacitor on the display board.
|
| |
|
Event or Problem Description
|
|
It was reported that the ar-8305 console is acting weird and had an electrical burning smell.
|
| |
|