|
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Failure mode: overheating insert.Root cause: no defect.Conclusion code: no failure found.Return qty.(b)(4), 30k thinsert-57, 00099052 bul.30k.Test method / gp005-wi02.Inductance: gauge id#: 5349, due: 11/30/2024, result 3.08.Meets spec, does not meet spec, n/a.Tip condition: meets spec, does not meet spec, n/a.Stack condition, meets spec, does not meet spec, n/a.Tested on: g130 gage id#: 9626-2, due date: 03/31/2024, set pressure: 35 psi.Water flow: output rate: 35 psi, meets spec, does not meet spec, n/a.Spray pattern: meets spec, does not meet spec, n/a.Water leak: hp sealing ring, proximal ring, distal ring.Seam leak, n/a.Vibration: meets spec, does not meet spec, n/a.Grip condition: meets spec, does not meet spec, n/a.Other visual observation: insert is good, no fault found.Insert was tested on a digital thermometer i.D.#: 6116a.Due date: 09/30/2024.The temperature was 83.3°f, no fault found.The specifications state the temperature is not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed, not confirmed.
|