DEPUY ORTHOPAEDICS INC US ARTIC/EZE TR BALL GRVD 28+1.5; FEMORAL TRIALS Back to Search Results
Catalog Number 253081000
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that surgeons requested new emphasys heads.It did not happen in surgery.Additional event information: visual analysis of the returned device found signs of worn out at the artic/eze tr ball grvd.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary it was reported that surgeons want new emphasys heads.It did not happen in surgery.The product was returned to depuy synthes for evaluation.Visual analysis of the returned device found signs of worn out at the artic/eze tr ball grvd.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the artic/eze tr ball grvd would have contributed to the device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (j0204) provided is not a valid finished goods lot# this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIC/EZE TR BALL GRVD 28+1.5
Common Device Name
FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19567170
Report Number1818910-2024-13273
Device Sequence Number6112737
Product Code HWT
UDI-Device Identifier10603295128946
UDI-Public10603295128946
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number253081000
Device Lot NumberJ0204
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2024
Initial Report FDA Received Date06/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2004
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-