Catalog Number 3501490
Medical Device Problem Codes Break (1069); Filling Problem (1233)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 05/13/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient who underwent breast augmentation revision with a mentor smooth round spectrum 575cc saline prosthesis experienced left sided difficulty with expansion.There was thought to be a kink and it created visibility of the device through the skin.As a result, explant was performed on (b)(6), 2024.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: difficulty with expansion / visible implant mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Additional information was received on june 24, 2024.In addition to the issues reported previously, the patient also experienced a filling tube issue.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The investigation of the photograph provided was completed by the failure analysis lab on july 23, 2024.Device evaluation summary: upon visual evaluation of the image provided in the complaint, tubing is observed kinked and cut.As the product involved in this complaint was not received, the device couldn´t be analyzed according to our procedures.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND SPECTRUM
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key19568015
Report Number1645337-2024-07388
Device Sequence Number6305894
Product Code FWM
UDI-Device Identifier00081317002048
UDI-Public00081317002048
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3501490
Device Lot Number9909364
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/30/2024
Supplement Date Received by Manufacturer06/24/2024
07/24/2024
Initial Report FDA Received Date06/19/2024
Supplement Report FDA Received Date07/16/2024
07/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-