| Model Number |
SPKD LGMNT STAPLE 11 MMX 20 MMLG |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023. arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint allegation cannot be confirmed without the device for evaluation.The device was not received for evaluation, and no photos of the failure were provided.Per the event description, the most likely cause(s) for the reported failure can be attributed to user error, improper bone preparation, misaligned insertion.
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl btb procedure, when the surgeon impacted the ar-1011, it twisted.A couple of attempts were made to reattach it but it would not work.The case was completed by using a new ar-1011.
|
| |
|