Model Number PEDICAP
Medical Device Problem Code Device Ingredient or Reagent Problem (2910)
Health Effect - Clinical Codes Unspecified Tissue Injury (4559); Unintended Extubation (4564)
Date of Event 03/07/2024
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, intra-operatively, there were two carbon dioxide detectors (co2 detectors) that did not change color after tracheal intubation.The co2 detector was placed at the end of the tube, and no color change was noted from purple to yellow.Another co2 detector was opened and placed again, and still no color change was noted.Multiple adjustments to the depth of the tube were made to see if it would correct the issue.This resulted in the infant being electively extubated despite clinical signs of intubation.It was also reported that the patient had serious injury.
 
Additional Manufacturer Narrative
Concomitant product: pedicap paed end tidal c02 detect (lot# 231630030).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX  22225
Manufacturer (Section G)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX   22225
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19570557
Report Number1282497-2024-00168
Device Sequence Number20382738
Product Code CCK
UDI-Device Identifier20884521083919
UDI-Public20884521083919
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K944400
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2025
Device Model NumberPEDICAP
Device Catalogue NumberPEDICAP
Device Lot Number231630030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2024
Initial Report FDA Received Date06/19/2024
Date Device Manufactured06/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-