| Model Number |
TROCAR FOR INFLOW CANNULA |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Event description: it was reported that there is rust on valve for inflow.It is inside the pipe and also visible when turning the handle.The reported event was partly confirmed as the visual evaluation revealed that the device has no valve or pipe.However the device shows signs of metal surface oxidation on the head (see evaluation picture 3).The most likely cause for the reported failure can be attributed to user error of the device due to improper detergent/cleaning process used.
|
| |
|
Event or Problem Description
|
|
It was reported that there is rust on valve for inflow.It is inside the pipe and also visible when turning the handle.There was no harm or adverse event for patient, operator or third party reported.No further information received.
|
| |
|
Search Alerts/Recalls
|