Model Number V+PRO, ENGLISH
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
An authorized third party service provider (asp) contacted ventec to report that the ventilator measured lower peep (positive end expiratory pressure) than set.There were no reports of patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
H6: the device was not returned to ventec for evaluation.The device was further evaluated by the authorized service provider (asp) where the reported issue of it measuring lower peep (positive end expiratory pressure) than set was confirmed.The asp replaced the internal flow transducer (ift) to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the ift.
 
Additional Manufacturer Narrative
Corrected information for supplemental medwatch report 001 -- section d4, model #, of the initial medwatch report stated: prt-01185-001.Section d4, model #, of the initial medwatch report should have stated: v+pro, english.Section d4, udi #, of the initial medwatch report stated: (b)(4).Section d4, udi #, of the initial medwatch report should have stated: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
2603 s goyer rd
kokomo IN 46902
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key19571185
Report Number3013095415-2024-00473
Device Sequence Number4160078
Product Code CBK
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K162877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation 505
Type of Report Initial,Followup
Report Date (Section B) 06/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberV+PRO, ENGLISH
Device Catalogue NumberPRT-01185-001
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2024
Supplement Date Received by Manufacturer03/24/2026
Initial Report FDA Received Date06/19/2024
Supplement Report FDA Received Date04/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/20/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-